Remote Control Commands
R&S
®
SMBV100A
672Operating Manual 1176.8016.02 ─ 17
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BB:DM:PRAM:BBON ON
selects power ramping in the baseband only.
Manual operation: See "Baseband Only" on page 371
[:SOURce<hw>]:BB:DM:PRAMp:FDELay <FDelay>
Sets the delay in the falling edge. A positive value gives rise to a delay and a negative
value causes an advance.
Parameters:
<FDelay> float
Range: -4 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example:
BB:DM:PRAM:FDEL 1
The falling edge starts 1 symbol later.
Manual operation: See "Fall Delay" on page 371
[:SOURce<hw>]:BB:DM:PRAMp:RDELay <RDelay>
Sets the delay in the rising edge. A positive value gives rise to a delay and a negative
value causes an advance.
Parameters:
<RDelay> float
Range: 0 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example:
BB:DM:PRAM:RDEL 1
The rising edge starts 1 symbol later.
Manual operation: See "Rise Delay" on page 371
[:SOURce<hw>]:BB:DM:PRAMp:SHAPe <Shape>
The command sets the edge shape of the ramp envelope.
Parameters:
<Shape> LINear | COSine
*RST: COSine
SOURce Subsystem