Instrument Function
R&S
®
SMBV100A
189Operating Manual 1176.8016.02 ─ 17
The remote commands required to define these settings are described in Chapter 7.5,
"BERT and BLER Subsystems", on page 532.
General Settings................................................................................................... 189
Configuration Settings...........................................................................................191
5.4.10.1 General Settings
State............................................................................................................................189
Bit Error Rate.............................................................................................................. 189
Data Bits......................................................................................................................189
Errors.......................................................................................................................... 189
Terminated By.............................................................................................................190
State:...........................................................................................................................190
State
Activates the bit error rate measurement. Depending on the selected Mode, you can
execute the measurement continuously ("Auto") or once ("Single"). To start a single
measurement, select "Execute Single".
Remote command:
:BERT:STATe on page 542
Bit Error Rate
Displays the measured bit error rate.
The bit error rate is the ratio of occurred errors to transmitted data bits.
The display unit can be selected in the box to the right of the value.
"ENG"
The bit error rate is output in exponential notation with the exponent
-3
or
-6
.
"SCI"
The bit error rate is output in scientific notation, i.e. standardized to
one place to the left of the decimal.
"%"
The bit error rate is output in percent.
"ppm"
The bit error rate is output in parts per million.
Remote command:
:BLER:RESult? on page 534
:BERT:UNIT on page 543
Data Bits
Displays the current number of checked data bits.
Remote command:
:BERT:RESult? on page 535
Errors
Displays the current number of occurred errors.
Remote command:
:BERT:RESult? on page 535
Bit and Block Error Rate Measurements