Remote Control Commands
R&S
®
SMBV100A
545Operating Manual 1176.8016.02 ─ 17
Suffix <hw>
Suffix Value range Description
CALibration<hw>
[1] Optional suffix
:CALibration:ALL[:MEASure]?......................................................................................... 545
:CALibration<hw>:BBIN[:MEASure]?................................................................................545
:CALibration:DATA:FACTory:DATE?................................................................................546
:CALibration<hw>:FMOFfset[:MEASure]?.........................................................................546
:CALibration<hw>:FREQuency[:MEASure]?......................................................................546
:CALibration<hw>:IQModulator:FULL?............................................................................. 546
:CALibration<hw>:IQModulator:LOCal?............................................................................547
:CALibration<hw>:LEVel:ALCTable[:MEASure]?............................................................... 547
:CALibration<hw>:LEVel[:MEASure]?...............................................................................547
:CALibration<hw>:LEVel:EXTern:DATA............................................................................548
:CALibration<hw>:LEVel:STATe...................................................................................... 548
:CALibration:ROSCillator[:DATA]..................................................................................... 548
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not require external measurement equipment.
Query parameters:
<Force> string
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CAL:ALL:MEAS?
// Response "0"
// Adjustment has been performed successfully
Usage: Query only
Manual operation: See "Adjust All" on page 121
:CALibration<hw>:BBIN[:MEASure]?
Starts adjustment of the analog I/Q input. The I/Q input is adjusted with respect to DC
offset and gain.
Return values:
<Measure> 0 | 1 | OFF | ON
*RST: 0
Example:
CAL:BBIN:MEAS?
starts the adjustment of the analog I/Q input.
Response: 0
adjustment has been performed successfully.
Usage: Query only