651Operating Manual 1176.8016.02 ─ 17
After selection, modulation parameters "Modulation Type" (:BB:DM:FORMat), "Symbol
Rate" (:BB:DM:SRATe), "Filter" (:BB:DM:FILTer:TYPE
and :BB:DM:FILTer:PARameter:...) and "Coding" (:BB:DM:CODing) are auto-
matically set in accordance with the standard.
The USER parameter cannot be set. A query returns this value if a user-defined Cus-
tom Dig Mod setting was loaded or if one of the associated settings was changed sub-
sequent to the selection of a standard. The user defined settings are stored and loaded
with commands :BB:DM:STAN:ULIS:... (see "DM Lists" on page 674).
Parameters:
<Standard> USER | BLUetooth | DECT | ETC | GSM | GSMEdge | NADC |
PDC | PHS | TETRa | W3GPp | TDSCdma | CFORward |
CREVerse | WORLdspace | TFTS | APCOPH1C4fm |
APCOPH1CQpsk | APCOPH2HCpm | APCOPH2HDQpsk |
APCOPH2HD8PSKW | APCOPH2HD8PSKN | APCOPH1Lsm |
APCOPH1Wcqpsk
*RST: GSM
Example:
BB:DM:STAN DECT
Selects digital modulation according to the DECT standard.
Manual operation: See "Set acc. to Standard" on page 355
[:SOURce<hw>]:BB:DM:STATe <State>
Enables/disables digital modulation. Switching on digital modulation turns off all the
other digital standards in the same signal path.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOURce1:BB:DM:STATe ON
Manual operation: See "State" on page 352
[:SOURce<hw>]:BB:DM:SWITching:SOURce <Source>
Selects the source of the CW control signal for switching between a modulated and an
unmodulated signal.
Parameters:
<Source> INTernal
INTernal
The CW signal in the control list is used for the control.
*RST: INTernal
Example:
BB:DM:SWIT INT
the CW signal in the control list is used for the control.
Manual operation: See "Source Mod - CW" on page 360
SOURce Subsystem