Remote Control Commands
R&S
®
SMBV100A
548Operating Manual 1176.8016.02 ─ 17
Return values:
<Measure> 0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See "Adjust Level" on page 121
:CALibration<hw>:LEVel:EXTern:DATA
<Data>
Queries what data has been used for the level calibration.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Parameters:
<Data> FACTory | CUSTomer
*RST: FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external
level correction.
Manual operation: See "Adjustment Data" on page 225
:CALibration<hw>:LEVel:STATe <State>
Activates internal level correction.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CAL:LEV:STAT
turns on level correction.
:CALibration:ROSCillator[:DATA] <Data>
Sets the calibration value for the custom defined external adjustment.
Parameters:
<Data> integer
Range: 0 to INT_MAX
*RST: 0
CALibration Subsystem