Remote Control Commands
R&S
®
SMBV100A
519Operating Manual 1176.8016.02 ─ 17
Example: 
{MARKER LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'MARKER LIST 1'
Queries the content of the MARKER LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
The marker setting for samples 0 to 9 = 0 (low), for 10 to 19 = 1
(high) and for 20 to 29 = 0. From sample 30 onward the marker
setting = 1.
Example: 
{LEVATT LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'LEVATT LIST 1'
Queries the content of the LEVATT LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
Level attenuation applies to data values 10 to 19 (high) and from
data value 30 onward.
Usage:  Setting only
{WAVEFORM-Length: #I0Q0I1Q1...IxQx...IN-1QN-1…}
(mandatory for waveforms)
The tag contains the actual waveform data or multi-segment waveform data (I/Q
stream).
Use the TYPE tag to define whether the waveform file contains a normal waveform or
a multi-segment waveform.
The following figure shows the principle of multi-segment work mode.
Waveform and List Format