Instrument Function
R&S
®
SMBV100A
131Operating Manual 1176.8016.02 ─ 17
2. In the "Setup > System > NRP-Z Info Update" dialog, select "Rescue" in the left
sensor selection field
3. Activate "Run Update".
4. Confirm query in message box
5. Connect sensor within 4 seconds
The update starts, a bar informs about the progress.
Current Sensors
Shows the sensors that are connected to the generator with information on serial num-
ber, the revision state and some features.
Remote command:
SENSe<ch>[:POWer]:TYPE? on page 588
SENSe<ch>[:POWer]:SVERsion? on page 587
SENSe<ch>[:POWer]:SNUMber? on page 586
Update
Section "Update" provides access to the file system in order to select a file for an R&S
NRP sensor update (Button "Select NRP File"), the selected file is indicated to the left
of the button. On the left side, the sensor to be updated is selected.
Button "Run Update" starts the update.
Note: If the update is interrupted for example by accidental removal of the sensor dur-
ing the process, the button "Rescue" appears. Thus, you can restart the update proc-
ess.
Prerequisite is that no other sensor is connected to the instrument.
Refer to "How to update an R&S NRP sensor" on page 130 and "How to restart an
interrupted update of an R&S NRP sensor" on page 130 for detailed instructions.
Remote command:
n.a.
5.2.3.6 Delete Temporary Files
During signal processing, the R&S SMBV creates and stores different temporary files,
such as the waveform files (*.wv) of all ARB standards. The instrument stores these
temporary files on a predefined directory on its internal memory and, if available, on
the installed hard disk. This predefined directory is a protected system directory that is
not accessible via the "File Manager" but you can delete all temporary files stored
there.
To achieve more free disc space on the internal flash memory
To maintain sufficient amount of free disc space on the internal flash memory or on the
hard disk, the R&S SMBV provides a convenient way to delete all temporary files
stored on the predefined directory.
1. Select "SETUP > Delete Temporary Files"
General Instrument Settings