Instrument Function
R&S
®
SMBV100A
240Operating Manual 1176.8016.02 ─ 17
Remote command:
[:SOURce<hw>]:CORRection:ZERoing:STATe on page 730
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
on page 725
Used SMBV Settings for Measurement
Displays the settings relevant for the measurement.
"RF Source"
Shows the path for which the correction menu settings are made.
"Modulation"
Indicates the modulation state
"Amplitude"
Shows the currently set level.
"Use SParameter"
Indicates whether SParameter correction is used.
"Attenuator Mode"
Displays the selected mode of the attenuator.
"Fixed Range (PEP) In:"
Shows the level range.
Remote command:
n.a.
5.5.10 Reverse Power Protection
The reverse power protection is tripped when the power of the external signal becomes
too high. A relay opens and interrupts the internal connection to the RF output. This
condition is indicated in the display header by the "OVERLOAD" status message.
Overload
If an "Overload" status message is indicated in the display header, reset the overload
protection by pressing the RF ON/OFF key.
The RF input is activated when the overload protection is reset.
Remote command:
:OUTPut<hw>:PROTection:TRIPped? on page 577
:OUTPut<hw>:PROTection:CLEar on page 577
:OUTPut<hw>[:STATe] on page 578
5.5.11 RF Measurement
5.5.11.1 NRP-Z Power Viewer
The R&S SMBV features the power viewer function for measuring or monitoring either
the RF output power, or a freely selectable signal source with R&S NRP-Z power sen-
sors.
The instrument can perform up to 4 power measurements simultaneously.
RF Block