Remote Control Commands
R&S
®
SMBV100A
547Operating Manual 1176.8016.02 ─ 17
Example:
CAL:IQM:FULL?
starts adjustments for the I/Q modulator across the entire fre-
quency range.
Response: "0"
adjustment has been performed successfully
Usage: Query only
Manual operation: See "Adjust I/Q Modulator Full Range" on page 121
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency, I/Q swap and
baseband gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q
imbalance and quadrature.
Return values:
<Local> 0 | 1
*RST: 0
Example:
CAL:IQM:LOC?
starts adjustment for the I/Q modulator for the currently set fre-
quency.
Response: 0
adjustment has been performed successfully
Usage: Query only
Manual operation: See "Adjust I/Q Modulator Current Frequency" on page 121
:CALibration<hw>:LEVel:ALCTable[:MEASure]?
Performs ALC adjustments according to the predefined values in the ALC table.
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CAL:LEV:ALCT:MEAS?
starts adjustments according to the ALC table.
Respone: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See " Adjust ALC Table - ALC " on page 228
:CALibration<hw>:LEVel[:MEASure]? [<Force>]
Starts all adjustments which affect the level.
Query parameters:
<Force> string
*RST: force
CALibration Subsystem