Remote Control Commands
R&S
®
SMBV100A
509Operating Manual 1176.8016.02 ─ 17
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code
Return values:
<ErrorCode> integer > 0 (in decimal format)
An error occurred.
(For details see the Service Manual supplied with the instru-
ment).
0
No errors occurred.
Usage: Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage:
Event
7.3 Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section. In addition, a specific preset command is provided for each
digital standard and for the fader. These specific commands are described in the asso-
ciated subsystems.
Four presetting actions are available:
Activating the default state of all internal instrument functions (*RST on page 507).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
Activating the preset state of the parameters related to the selected signal path (:
SOURce<hw>:PRESet on page 510)
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet on page 509)
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 510). Only functions that are protected by a password remain
unchanged as well as the passwords themselves.
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Preset Commands