Remote Control Commands
R&S
®
SMBV100A
549Operating Manual 1176.8016.02 ─ 17
7.7 CLOCk Subsystem
The CLOCk system contains the commands for configuration of the signals at the clock
output and input connectors.
:CLOCk:INPut:FREQuency?........................................................................................... 549
:CLOCk:INPut:SLOPe.................................................................................................... 549
:CLOCk:INPut:FREQuency?
Queries the measured frequency of the external clock signal. An external clock refer-
ence must be supplied at the CLOCK input.
Return values:
<Frequency> float
Range: 0 to max
Increment: 0.001
*RST: 0
Example:
CLOC:INP:FREQ?
queries the measured frequency of the external clock reference.
Usage: Query only
Manual operation: See "Measured External Clock" on page 378
:CLOCk:INPut:SLOPe <Slope>
Sets the active slope of an externally applied clock signal at the CLOCK connector.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example:
CLOC:INP:SLOP NEG
The active slope of the external clock signal is the falling slope.
Manual operation: See "Clock Input Slope" on page 141
7.8 DIAGnostic Subsystem
The DIAGnostic system contains the commands used for instrument diagnosis and
servicing. SCPI does not define any DIAGnostic commands; the commands listed here
are all device-specific. All DIAGnostic commands are query commands which are
not influenced by *RST.
:DIAGnostic<hw>:BGINfo?..............................................................................................550
:DIAGnostic<hw>:BGINfo:CATalog?................................................................................ 550
:DIAGnostic:INFO:OTIMe?..............................................................................................551
:DIAGnostic:INFO:POCount?.......................................................................................... 551
DIAGnostic Subsystem