Remote Control Commands
R&S
®
SMBV100A
581Operating Manual 1176.8016.02 ─ 17
Example:
SENS:UNIT DBM
selects unit dBm for presentation of measurement result.
READ1?
queries the measurement result of the sensor connected to the
SENSOR interface.
Response: -45.6246576745440230
-45.6 dBm were measured at the given frequency.
or e.g. for R&S NRP-Z81
Response:
-55.62403263352178,-22.419472478812476
-55,6 dbm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency
Usage:
Query only
Manual operation: See " Level (Peak) " on page 244
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
<Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select> float
*RST: 0
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe <State>
The command activates the use of the s-parameters correction data of the selected
power sensor.
Note: For power sensor with attenuator this command is automatically set to ON.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
SENS:POW:CORR:SPD:STAT ON
activates the use of the s-parameters correction data of power
sensor 1.
Manual operation: See " Use SParameter - Power Sensors " on page 232
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
Return values:
<List> string list
*RST: 0
SENSe, READ and INITiate Subsystems