Remote Control Commands
R&S
®
SMBV100A
582Operating Manual 1176.8016.02 ─ 17
Usage: Query only
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe <State>
The command switches on and off the permanent indication of the power measure-
ment result in the upper right corner of the block diagram. For each sensor, the type of
sensor, the connector, the measurement source and - if set - the offset is indicated.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: ON
Example:
SENS1:POW:DISP:PERM:STAT ON
the permanent viewer is switched on.
Manual operation: See " Permanent Display State " on page 244
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority
<Priority>
The command selects which power measurement result (average or peak power) is
indicated when permanent display is active.
Parameters:
<Priority> AVERage | PEAK
*RST: PEAK
Example:
SENS1:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
the measured average power is indicated.
Manual operation: See " Display Priority " on page 245
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE AUTO)
Return values:
<Auto> float
Range: 1 to 65536
Example:
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Usage: Query only
Manual operation: See "Filter Length" on page 247
SENSe, READ and INITiate Subsystems