Trying out the Instrument
R&S
®
SMB100A
57Quick Start Guide 1407.0812.62 ─ 20
To indicate the active state, the RF block is displayed in blue. An RF signal with the
default frequency and level settings is output, i.e. 1 GHz and -30 dBm.
The sweep is not yet active, it must be triggered in the sweep dialog.
4. Trigger RF frequency sweep
a) Press the [Winbar] key to switch to the "RF Frequency Sweep" dialog.
Turn the rotary knob to select the "Execute Single Sweep" button.
Press the rotary knob to trigger (start) the frequency sweep.
A linear single sweep signal is now present at the RF output, starting at 100 MHz.
The sweep is processed in 1 MHz steps with dwell time of 10 ms per step up to the
stop frequency of 500 MHz.
The sweep starts at 100 MHz, stops at 500 MHz in 1 MHz steps is output with a
dwell time of 10 ms per step.