Command Reference
R&S
®
ZNB/ZNBT
1009User Manual 1173.9163.02 ─ 54
Parameters:
<Logic> POSitive | NEGative
POSitive 0 = low, 1 = high
NEGative 0 = high, 1 = low
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
*RST: n/a (default: NEGative)
Example:
CONTrol:HANDler:LOGic POS
Change the logic of the data ports to positive.
CONTrol:HANDler:OUTPut<Pt>:USER <BinValue>
Defines the state of the output ports (pin 3 or 4) of the Universal Interface connector
after the next negative pulse on the /INPUT1 line (pin 2).
Suffix:
<Pt>
.
Output port number:
1 - /OUTPUT1 (pin 3)
2 - /OUTPUT2 (pin 4)
The parameters MIN, MAX, DEF are not supported.
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
Parameters:
<BinValue> 0 - high 1 - low
*RST: n/a (default: 0)
Example: See CONTrol:HANDler:OUTPut<Pt>[:DATA]
CONTrol:HANDler:OUTPut<Pt>[:DATA] <BinValue>
Writes a 0 or 1 to the output ports (pin 3 or 4) of the Universal Interface connector. The
port lines have negative logic: A "0" corresponds to a high signal, a "1" to a low signal.
The query reads the last value that has been written to the output port.
Suffix:
<Pt>
.
Output port number:
1 - /OUTPUT1 (pin 3)
2 - /OUTPUT2 (pin 4)
The parameters MIN, MAX, DEF are not supported.
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
SCPI Command Reference