Command Reference
R&S
®
ZNB/ZNBT
1126User Manual 1173.9163.02 ─ 54
<L0>, <L1>,
<L2>, <L3>
mandatory* mandatory mandatory* mandatory** mandatory**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional*** optional*** optional*** mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1> optional: <Port 1>,<Port 2>
* values will be ignored during calibration
** if OPEN is selected the residual inductance (L0,...,L3) will be ignored during calibration; if SHORt is selected the fringing capaci-
tance (C0,...,C3) will be ignored during calibration
*** must be provided if a port restriction <Port 1> shall be applied; values will be ignored during calibration
The parameters in the [SENSe<Ch>:]CORRection:CKIT:<StandardType>,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType> com-
mands have the following meaning:
Table 8-13: Parameter list
Parameter Meaning Comment/Unit
'<ConnType>' Name of the connector type.
Use SENSe1:CORRection:CONNection:CATalog? to query connec-
tor names.
String parameter
'<CalKitName>' Name of the calibration kit. Use
[SENSe:]CORRection:CKIT:CATalog? to query cal kit names.
String parameter
'<CalKitLabel>' Label (e.g. the serial number) of the calibration kit; for ...WLABel com-
mand only
String parameter
'<StandardLa-
bel>'
Label (e.g. the serial number) of the standard String parameter
<MinFreq>,
<MaxFreq>
Min./max. frequency for which the circuit model is valid Default unit is Hz
<DelayParam>
Depending on the cal kit's model type (selected using [SENSe:]CORRection:CKIT:DMODe) this parameter is
interpreted as
delay [s] for Agilent modelling
el. length [m] for ZVR compatible modelling
Cal kits that are not created/modified using [SENSe:]CORRection:CKIT:DMODe use the ZVR compatible
modelling.
<Loss> Loss (offset parameter) of the standard To be specified without unit (implicit
unit is dB)
<Z0> Reference impedance (no unit) To be specified without unit (implicit
unit is Ω)
<C0>,...,<C3> Polynomial coefficients for the fringing capacitance of the standard (load
parameter)
To be specified without unit: implicit
unit of <Ci> is fF / (GHz)
i
<L0>,...,<L3> Polynomial coefficients for the residual inductance of the standard (load
parameter)
To be specified without unit: implicit
unit of <Li> is pH / (GHz)
i
SCPI Command Reference