Command Reference
R&S
®
ZNB/ZNBT
1178User Manual 1173.9163.02 ─ 54
Parameters:
<Boolean> ON (1): compensation active
OFF (0): compensation inactive
Manual operation: See "Active" on page 572
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Returns whether a direct fixture compensation has been carried out at port no.
<PhyPt>. A direct fixture compensation resets the offset parameters to zero, the ana-
lyzer uses calculated transmission factors instead.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer.
Return values:
<Boolean> 1 - direct fixture compensation data used
0 - no direct fixture compensation data used
Example:
*RST; CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct
fixture compensation data at port 1. The response is 0.
Usage: Query only
Manual operation: See "Direct Compensation" on page 585
[SENSe<Ch>:]CORRection:PCAL <ActivatePowerCals>
Activates or deactivates all power calibrations.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<ActivatePowerCals> NONE | ALL
NONE deactivate all power calibrations.
ALL activate all power calibrations.
*RST: n/a
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is 1).
CORR:PCAL NONE; POW:AWAV?
Disable all power calibrations and check again whether the cali-
bration is applied (the response is 0).
Usage: Setting only