Command Reference
R&S
®
ZNB/ZNBT
1246User Manual 1173.9163.02 ─ 54
-7, "Swept mode not possible while AVG Detector is used."
-8, "Swept mode not possible: Lin Frequency Sweep only supported!"
-9, "Swept mode not possible while Arbitrary or Intermodulation Measurement is
used."
-10,"Swept mode not possible: this combination of features is not yet supported!"
-11, "Swept mode not supported. Please contact Rohde&Schwarz Service for an
upgrade."
This may occur for some "old" synthesizer hardware.
Suffix:
<Ch>
.
Channel number
Return values:
<arg0> Status code
Negative values indicate that swept mode could not be activated
for the reason described in the <state_desc>.
<arg1> State description
Usage: Query only
Manual operation: See "Freq Sweep Mode" on page 395
[SENSe<Ch>:]SWEep:POINts <SweepPoint>
Defines the total number of measurement points per sweep ("Number of Points").
Values between 1 and 100,001 can be set.
Suffix:
<Ch>
.
Channel number 1
Parameters:
<SweepPoint> Number of points per sweep
Range: see above
*RST: 201
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation: See "Number of Points" on page 384
SCPI Command Reference