Command Reference
R&S
®
ZNB/ZNBT
1245User Manual 1173.9163.02 ─ 54
[SENSe<Ch>:]SWEep:DWELl:IPOint <InsertionPoints>
Defines whether the measurement delay (previously defined via [SENSe<Ch>:
]SWEep:DWELl) is inserted before all partial measurements or before the first partial
measurement only.
Suffix:
<Ch>
.
Channel number
Parameters:
<InsertionPoints> ALL | FIRSt
Insertion before all or before the first partial measurement
*RST: ALL
Example: See [SENSe<Ch>:]SWEep:DWELl
Manual operation: See "All Partial Meas'ments / First Partial Meas'ment"
on page 395
[SENSe<Ch>:]SWEep:GENeration <arg0>
Sets/gets the sweep mode for linear frequency sweeps.
For segmented sweeps this can be overridden by segment-specific sweep modes
([SENSe<Ch>:]SEGMent<Seg>:SWEep:GENeration).
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> STEPped | ANALog
Stepped mode or swept mode (=ANALog).
*RST: STEPped
Manual operation: See "Freq Sweep Mode" on page 395
[SENSe<Ch>:]SWEep:GENeration:ANALog:CONDition?
Returns the state of the swept mode (see [SENSe<Ch>:]SWEep:GENeration and
[SENSe<Ch>:]SEGMent<Seg>:SWEep:GENeration), which can be one of the fol-
lowing:
0, "Swept mode not active."
1, "Swept mode active."
(for at least one segment in segmented sweeps)
-1, "Swept mode not possible while external powermeter is used."
-2, "Swept mode not possible while external generator is used."
-3, "Swept mode not possible while Point or Partial Measurement Trigger is used."
-4, "Swept mode not possible while Point Delay is used."
-5, "Swept mode not possible while Chopped Driving Mode is used."
-6, "Swept mode not possible while DC Measurement is used."