Firmware V3.15
R&S
®
ZNB/ZNBT
19User Manual 1173.9163.02 ─ 54
●
Formula-defined limit lines: formula was not applied if entered directly in
"Response" dialog
●
Logarithmic interpolation of limit lines did not work for linear sweeps
●
Missing space between cal label and power cal label in trace info
●
Error message for missing calibration not informative
●
Installation of R&S ZNB/BT/D firmware simulation failed if R&S ZVA/B/T firmware
simulation was installed on the same PC
●
When storing calibration files using MMEM:STOR:CORR, file names containing dots
were truncated
●
Segmented sweeps: in FW versions 3.x, markers could not be positioned between
sweep segments
●
PC firmware simulation failed to run due to incorrect environment detection
●
Fixed an issue with handling multiple Line standards in TRL calibration
Firmware version
► To check your R&S ZNB/ZNBT firmware version, select "Help" > "About..." from the
main menu.