Firmware V3.15
R&S
®
ZNB/ZNBT
18User Manual 1173.9163.02 ─ 54
Improvements
"Distance to Fault" measurements (R&S ZNB-K3/R&S ZNBT-K3):
Import dialog for cable types
See "Import Cable Type" on page 682
Now available on all ports
Now also available in "Applic" softtool
See Chapter 6.15.4, "Distance to Fault Application", on page 676
Optimized calculation of port set de-/embedding
Extended time domain analysis (option R&S ZNB/ZNBT-K20): new "TDR Wizard"
to set up and calibrate a TDR measurement quickly (R&S ZNB only!)
See Chapter 6.15.2.2, "TDR Wizard", on page 633
Enhanced selftest functionality:
Detailed self test report now also available at service level 0
Selftest execution can be triggered directly from SYSTEM – [SETUP] > "Setup"
softtool tab
Test report comprises serial number and firmware version
Overall and individual test results clearly highlighted ("PASS": green, "FAIL":
red)
See "Selftest Tab" on page 715
Auto length and loss functions:
Approximate loss can be calculated based on loss values at two positive refer-
ence frequencies instead of one
Curve fitting can be restricted to a custom "Evaluation Range"
Improved handling of long diagram titles during printing
Configurable system defaults for trace-specific USER PORT "TTL1 Pass" and
"TTL2 Pass" settings
See "TTL Pass Default Values" on page 709
Support of R&S NRQ6 power sensor via LAN
Solved issues
Segmented frequency sweeps:
If the column "Segm Time" was not displayed in the "Define Segments" dialog, the
total sweep time ([Sweep] > "Sweep Params" > "Sweep Time"
was not calculated from the segment times
was configurable
Deleting a trace by dragging it to the recycle bin sometimes did not work
Configured "Reference Receiver Cal Power" value not used for SMARTerCal
R&S ZNBT segmented sweeps: AGC "Auto" mode in one segment activated the
AGC pre-measurement for measurements in all segments
Multiple DUT configurations: for segmented sweeps with segment-specific power,
the AGC learn sweep did not work
Calibrations using cal unit R&S ZN-Z150 resulted in an error message