Measure
Starts the measurement of DUT + Test Fixture at the Active ports.
The display will change to show all the S parameter measurements being made. The
resulting Touchstone file is written to C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Embedding.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure
CALCulate:FMODel:SFD<Ph_pt>:DUT:MEASure
Active
Before starting to Measure, use these checkboxes to indicate the physical ports to
which DUT + test fixture are connected. For single-ended deembedding 2 ports must
be active, for balanced deembedding 4 ports.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe]
CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]
6.13.5.4 Save Fixture Model
This section allows you to run the selected deembedding tool to generate the Touch-
stone files for fixture deembedding (and the Touchstone file of the de-embedded DUT).
Figure 6-55: Save Fixture Model section
Offset Embed Softtool