GUI Reference
R&S
®
ZNB/ZNBT
600User Manual 1173.9163.02 ─ 54
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:SCALe:FTIMe
Automatic Flt Tm for DUT + Lead Ins ← Lead Ins
If checked (default) the flight time for DUT + lead-ins is calculated automatically. Other-
wise it can be specified (see "Flt Tm for DUT + Lead Ins" on page 600).
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride
Flt Tm for DUT + Lead Ins ← Lead Ins
If Automatic Flt Tm for DUT + Lead Ins is disabled, this allows you to set the flight time
for DUT + Lead-ins manually.
This should be used if the DUT + Test Fixture does not have a through response and
the total flight time exceeds 10ns.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:DUT
DUT With Fixture
Defines how the ISD tool shall interpret the DUT + Fixture data.
Port Sequence ← DUT With Fixture
Tells the ISD tool about the arrangement of DUT + Fixture ports:
"1 to N on left": ports 1 to N are on the left and ports N+1 to 2*N are on the right.
"Odd on left" (default): 1, 3, 5, etc. are on the left and ports 2, 4, 6, etc. are on the
right.
"All on left": all ports are on the left
Note: Ports on the left/right are assumed to be coupled (if Trace Coupling isn't set to
"None". A port on the left side is not considered to be coupled to a port on the right
side.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer
DUT Type ← DUT With Fixture
Select "Passive"/"Active" if the DUT is passive/active.
Default is "Passive".
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:DUT:TYPE
Ports to Skip ← DUT With Fixture
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run
"None" (default): do not skip any ports
"Ports on left": skip the ports on the left (according to the selected Port Sequence)
"Ports on right": skip the ports on the right (according to the selected port
sequence)
"Manually set ports": use Ports to Skip (manual) to define the ports to be skipped
Offset Embed Softtool