615User Manual 1173.9163.02 ─ 54
6.13.11.1 Delta-L Settings Dialog
From the "Detal-L Settings" dialog, you can set up the Delta-L measurement channel
and how the external tool is run.
Access: CHANNEL – [OFFSET EMBED] > "Delta-L" > "Delta-L Settings..."
Method
Selects the Delta-L+ characterization method to be used. For background information,
see Delta-L 4.0 PCB Characterization.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:METHod
Measurements
"Method 1L" handles single-ended and balanced measurement results differently. If
this Method is used, you have to select one of them.
"Method 2L" and "Method 3L" measurements can also be single-ended or balanced,
but the tool does not distinguish between them.
Remote command:
CALCulate:FMODel:DELT<Ph_pt>:MEASurement
Port Order
Declares how the DUT is connected to the measurement system. Choosing the wrong
port ordering for your setup, can result in erroneous results.
"Sequential"
Ports 1, 2 and 3, 4 form input/output pairs
"Odd / Even"
Ports 1, 3 and 2, 4 form input/output pairs
Offset Embed Softtool