Command Reference
R&S
®
ZNB/ZNBT
794User Manual 1173.9163.02 ─ 54
Suffix:
<Chn>
.
Channel number used to identify the active trace
Return values:
<Boolean> 0 – skew check has passed
1 – skew check has failed
Usage: Query only
Options: R&S ZNB/ZNBT-K20
CALCulate<Chn>:DTIMe:LIMit:FAIL:BEEP <Boolean>
Defines whether the R&S ZNB/ZNBT should make an audible beep on skew limit viola-
tions.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Skew Fail Beep" on page 659
CALCulate<Chn>:DTIMe:LIMit:LIMit <arg0>
Defines the limit value for the skew check.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<arg0> Limit value in seconds or meters, depending on the format of the
current trace's stimulus axis (see CALCulate<Chn>:
TRANsform:TIME:XAXis).
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Skew Limit" on page 659
CALCulate<Chn>:DTIMe:LIMit:STATe <Boolean>
Activates/deactivates the skew limit check.
Use CALCulate<Chn>:DTIMe:LIMit:LIMit to set the applicable limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNB/ZNBT-K20
SCPI Command Reference