Loading...
Command Reference
R&S
®
ZNB/ZNBT
852User Manual 1173.9163.02 ─ 54
8.3.1.9 CALCulate:GDAPerture...
The CALCulate:GDAPerture... commands configure the group delay measure-
ment.
CALCulate<Chn>:GDAPerture:SCOunt <Steps>
Defines an aperture for the calculation of the group delay as an integer number of fre-
quency sweep steps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Steps> Number of steps
Range: 1 to 10000
*RST: 10
Example:
*RST; :CALC:FORM GDEL
Select group delay calculation for the active trace.
CALC:GDAP:SCO 15
Select an aperture of 15 steps.
Manual operation: See "Aperture Points" on page 296
8.3.1.10 CALCulate:IAVerage
CALCulate<Chn>:IAVerage[:STATe] <Boolean>
Turns infinite averaging on or off.
The averaging mode can be selected using CALCulate<Chn>:IAVerage:MODE.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Infinite Average/Restart" on page 335
CALCulate<Chn>:IAVerage:MODE <Mode>
Selects the quantities to be averaged if infinite averaging is enabled
(CALCulate<Chn>:IAVerage[:STATe] ON.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Mode> MAGPhase | REIMag
Selects the averaging mode.