Command Reference
R&S
®
ZNB/ZNBT
868User Manual 1173.9163.02 ─ 54
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
Manual operation: See "Segment List" on page 348
CALCulate<Chn>:LIMit:SOUNd[:STATe] <Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Fail beep on or off.
*RST: OFF
Example:
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation: See "Limit Fail Beep" on page 345
CALCulate<Chn>:LIMit:STATe <Boolean>
Switches the limit check (including upper and lower limits) on or off.
Tip: Use CALCulate<Ch>:LIMit:UPPer:STATe or
CALCulate<Ch>:LIMit:LOWer:STATe to switch on or off the individual limit checks
for upper or lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Limit check on or off.
*RST: OFF
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
Manual operation: See "Limit Check" on page 344
CALCulate<Chn>:LIMit:STATe:AREA <HorizontalPos>, <VerticalPos>
Moves the limit check pass/fail message for the active trace <Chn> to one of nine pre-
defined positions in the active diagram.
SCPI Command Reference