Command Reference
R&S
®
ZNB/ZNBT
917User Manual 1173.9163.02 ─ 54
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:RIPPle:FAIL?
Manual operation: See "Ripple Check" on page 352
8.3.1.17 CALCulate:SMOothing...
The CALCulate:SMOothing... commands provide the settings for trace smoothing.
CALCulate<Chn>:SMOothing[:STATe]..............................................................................917
CALCulate<Chn>:SMOothing:APERture.......................................................................... 917
CALCulate<Chn>:SMOothing[:STATe] <Boolean>
Enables or disables smoothing for trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - smoothing on or off.
*RST: OFF
Example: See CALCulate<Chn>:SMOothing:APERture
Manual operation: See "Smoothing" on page 332
CALCulate<Chn>:SMOothing:APERture <SmoothAperture>
Defines how many measurement points are averaged to smooth the trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SmoothAperture> Smoothing aperture. A smoothing aperture of n % means that
the smoothing interval for each sweep point i with stimulus value
x
i
is equal to [x
i
- span*n/200, x
i
+ span*n/200], and that the
result of i is replaced by the arithmetic mean value of all mea-
surement points in this interval.
Range: 0.05% to 100%.
*RST: 1
Default unit: %