2.7 M1S- & Roline-LxS Series connected to PC via MOK-WIN and ACRLXB5
Device
M1xS
M1xS
L1xS
Possible
Functions
Accessories required
Yes Yes
Display the measurement
values of the M1 / L1x
transmitter
Yes Yes
1-point adjustment using a
RHS humidity standard.
Yes Yes
Multi- point adjustment using
a RHS humidity standard.
Yes Yes
1 point or 2 point adjustment
of temperature
Yes Yes
Display the serial number, last
calibration date of
temperature and humidity
calibration)
PC running
HW3 software
AC1625
MOK-02-WIN
(eventually
AC adaptor AC1207)
Service cable
MOK-02-WIN
Note:
HW3 will recognize the transmitters as a single HygroClip probe.
Do not attempt to adjust temperature unless you have very accurate reference
instruments!
Note:
The shaded functions can be performed with the distributor version of HW3 only.
\\Kalibrierung\Calibration_procedures_2.doc