SaeboStim Pro
Care and Maintenance
Device Storage
• Do not store in direct sunlight, high temperature, moist, dusty, or corrosive
gas.
• Store where children cannot reach.
• Use a wet cloth with mild detergent or alcohol to clean the surface of the
device.
• Do not immerse the electronic components into water.
Electrodes Maintenance
• Keep the gel surface of the Electrode Pad clean.
• AfterusingtheElectrodePad,usethelmcovertoprotectit.
• When the Electrode Pad is dirty or when the skin is tingling, rinse with a
smallamountofwaterandgentlyscrubwithyourngersforafewseconds.
After drying, it can temporarily restore its viscosity. If the water is too much,
the viscosity will decrease.
• Please do not use tissue or cloth to wipe the gel surface.
• Pleasedonotusengernailorbrushtoscrapethegelsurface.
Replace Batteries
• The battery icon will appear at all times during operation in the top center of the
display.
• When the stimulator is in use, the individual bars on the battery icon disappear
sequentially.Whenthebatteryhasonebarleft,theiconwillashtoindicatethe
replacement of batteries or power via DC jack.
• Once all the bars have disappeared, the stimulator will beep for 10 seconds,
batteryiconwillash,andthestimulatorwillstopandpowero.
39