2. Installation
19
2.1 Installation
2.1.1 General Information on Installing the Instrument
This section provides information on determining a location for and installing
the instrument.
Installation should be performed by trained staff for the instrument. This
instrument must be installed correctly to ensure proper operation and service.
Read this Operating Manual carefully before operating the instrument. Follow
all instructions provided in this document carefully.
CAUTION:
This instrument is a precision instrument and must be
handled with care. If the instrument is handled in a rough
manner or dropped, its internal components may malfunction
or get damaged. Always exercise due care when handling the
instrument.
2.1.2 Installation Environment
Install the instrument in a location meeting the following conditions:
Free from re or ignition sources nearby
Not exposed to direct sunlight
A horizontal level and stable surface area
Subject to minimum vibration
Protected from water splashes
Away from volatile substances or ammable materials
Away from corrosive chemicals (acids, alkalis, etc.)
Free from equipment that consumes high voltage or large current
Example) Large refrigerator, ultrasonic cleaner, cooling centrifuge, etc.
The ambient temperature and humidity meet the following use
environment values:
Ambient temperature: 10 to 35°C
Relative humidity: 30 to 85 % (non-condensing)
CAUTION