28
2.2.2 Setting the Schedule of the Auto Operation (with the
Auto Cryo Mode)
This instrument is operated in the auto mode and manual mode.
In the auto mode, the instrument is automatically activated so that paraffin
has melted sufciently when the start time arrives. When the end time arrives,
the instrument automatically pauses.
The following explains how to set the days of the week when auto operation
is performed, and the start and end times of the auto operation.
NOTE:
Set the current date/time correctly before auto operation can be
performed.
1. Touch the “Main Menu” button [1] on
the 2.Operation screen.
2. Touch the “Auto” button [2] on the
Main Menu screen.
3. The Auto Operation Display screen is
displayed.
The start and end times of operation
on Sunday through Saturday are
displayed.
Touch "
"/" " on the screen to
switch between the display of the
rst half of the week and the second
half of the week.
4. Auto operation can be set separately
for each day of the week.
A “
•” on the left side of the day-of-
the-week field indicates that auto
operation is turned on for that day of
the week.
A “
•” indicates that auto operation is turned off.
Touch the left side of the day-of-the-week field to switch the auto
operation setting between on and off.
[1]
[2]
* The page showing the first half of the week
displays Sunday through Wednesday, while the
page showing the second half of the week displays
Thursday through Saturday.