24
4. Insert the power plug of each module into an electrical outlet with
grounding terminals.
Power supply system meeting the following conditions is needed to
connect the power plugs:
Voltage: 100VAC 115VAC 230VAC
Capacity: EM 15 A or more 15 A or more 10 A or more
CM 15 A or more 15 A or more 10 A or more
CAUTION:
Connect the instrument power cord directly to an electrical
outlet instead of extending it using a different cord, etc.
Especially in the case of the Cryo Module, if the power cord
is extended and connected, the supply voltage may drop and
the compressor may stop working.
CAUTIONS:
• Do not connect/remove the power plug with a wet hand.
Failure to observe the instruction may result in electric shock.
When connecting/removing the power plug, do so by holding
the plug. Connecting/removing the power plug by holding
parts other than the plug may cause fire or electric shock.
• Connect the power plug to a dedicated power outlet with
grounding terminals. If the power plug is not grounded
through proper grounding terminals, electric shock may
result.
5. Determine the position of the instrument. (The rear and sides of the
instrument may be in contact with walls, etc.)
6. By trainned staff, an offset value for temperature compensation, which is
labeled on the backside of the Cryo Module, is entered in the Embbeding
Module.