2. Installation
31
Details of the Auto Operation Mode
Cryo Module
Auto operation of the Cryo Module is performed with the Auto Cryo mode
on or off, as selected.
* Auto Cryo mode on:
The Cyro Module activates/pauses automatically in conjunction with the
auto operation of the Embedding Module. During the time period when
the Cryo Module is operating, the module can also be activated/paused
using the “Cryo” button.
* Auto Cryo mode off:
The user touches the “Cryo” button to switch the Cryo Module state
between active and paused.
When the Embedding Module is paused, the Cryo Module is also
paused automatically.
30 minutes
OperatingPaused
0 AM 0 AM 0 AM
Day 1 Following day
TEC 6-EM
start time 1
TEC 6-EM
end time 1
TEC 6-EM
start time 2
TEC 6-EM
end time 2
Paused
Instrument
state
TEC 6-CM
Operating Paused
30 minutes