9 - 26 RS80A Service Manual
Disinfection
Disinfect the probe using a disinfectant solution recommended by Samsung
Medison to reduce pathogens to the level of 10
-6
.
WARNING
If a pre-mixed solution is used, be sure to observe the
solution expiration date.
The type of tissue it will contact during use dictates the level
of disinfection required for a probe. Ensure that the solution
strength and duration of contact are appropriate for
disinfection.
CAUTION
Failing to use a recommended disinfectant or follow the
recommended disinfection method may damage and/or
discolor the probe. This also voids the product warranty.
Do not immerse non-sterilizable probes in disinfectant for
more than one hour.
Only use liquid solutions to sterilize probes. Do not sterilize
by autoclaving or using EtO gas.
1. Please refer to the operation manual of the disinfectant for details of proper
storage, use, and disposal of the disinfectant.
2. Mix the disinfectant compatible with your probe according to label
instructions for solution strength..
3. Immerse the probe into the disinfectant as shown in the illustration below.
4. Follow the instructions on the disinfectant to complete the immersion
procedure of the probe, and rinse the probe afterwards.
5. Allow the probe to air dry or wipe it with a dry cloth.