2 - 14 RS80A Service Manual
CAUTION
If electromagnetic waves are causing interference with the
operation of the system, relocate the product far away from the
EMI.
EMC
The testing for EMC (Electro-Magnetic Compatibility) of this system has been
performed according to the international standard for EMC with medical devices
(IEC-60601-1-2). In Europe, the European Standard (EN 60601-1-2) is adopted as
the IEC Standard.
▐ Guidance and Manufacturer’s Declaration –
Electromagnetic Emission
This product is intended for use in the electromagnetic environment specified
below. The user must make sure that this product is used in the following
environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF emission
CISPR 11
Group 1
The ultrasound system only uses RF energy to
operate the internal functions. Therefore, the
RF energy emitted is very low, and unlikely to
cause interference to nearby electronic
equipment.
RF emission
CISPR 11
Class A
The ultrasound system is suitable for use in all
environment except residential environment.
By paying attention to the following warning,
the system may be used in environments
where electricity is supplied to a residential
Harmonic emission
IEC 61000-3-2
Class A