Chapter 3. Installing the Product 3-9
How to install the Product
Installation Safety
Power receptacle and equipotential terminal are located at the rear of product.
[Figure 3.2 Power Receptacle and Equipotential Terminal]
Power inlet: Accepts the power cord, which connects to an external power
supply.
Equipotential terminal: Must be connected to the equipotential bonding in a
treatment room.