Chapter 2. Safety 2 - 17
Immunity Test
IEC 60601
Test Level
Regulation Level
Electromagnetic
Environment -
Guidance
Voltage dips,
short
interruptions, or
voltage variations
in power supply
cables
IEC 61000-4-11
<5%Uт within 0.5
cycle
(>95% dip, unit:
Uт)
40%Uт within 5
cycles
(60% dip, unit: Uт)
70%Uт within 25
cycles
(30% dip, unit: Uт)
<5%Uт for 5
seconds
(<95% dip, unit:
Uт)
<5%Uт within 0.5
cycle
(>95% dip, unit:
Uт)
40%Uт within 5
cycles
(60% dip, unit: Uт)
70%Uт within 25
cycles
(30% dip, unit: Uт)
<5%Uт for 5
seconds
(<95% dip, unit:
Uт)
The quality of main
power supply must be
equal to the quality of
power supply that is used
in general commercial or
medical environments. If
operation must continue
during a failure of the
main power, it is
recommended that power
is supplied through an
uninterruptible power
supply or a battery .
Commercial
frequency
(50/60Hz) EM
field
IEC 61000-4-8
3A/m 3A/m
The electromagnetic field
from power frequency
must be equal to the EM
field in general
commercial or medical
environments.
Note: Uт is the primary voltage (AC) before applying test level.