- 25 -
6-9
Indication of Message
Error codes
Indication
Meaning
“Error 01: Temperature Sensor Open” Temp. sensor is open circuited.
“Error 02: Temperature Sensor Short” Temp. sensor is short circuited.
“Error 03: Condenser Sensor ‘A Open” Sensor at exit of condenser in A circuit is open circuited.
“Error 04: Condenser Sensor ‘A Short” Sensor at exit of condenser in A circuit is short circuited.
“Error 05: Condenser Sensor ‘B’ Open Sensor at exit of condenser in B circuit is open circuited.
“Error 06: Condenser Sensor ‘B’ Short” Sensor at exit of condenser in B circuit is short circuited.
“Error 07: Ambient temp Sensor Open” AT sensor is open circuited.
“Error 08: Ambient temp Sensor Short” AT sensor is short circuited.
“Error 09: Battery switch is off” Battery switch is in off position.
“Error 10: Condenser temp is abnormal” Condenser temp. is abnormal. (Both fan motor are
locked )
“Error 16: Fan motor ‘Aabnormal” Condensing fan motor A is abnormal.(Condensing fan
motor A is locked)
“Error 17: Fan motor ‘B’ abnormal” Condensing fan motor B is abnormal.(Condensing fan
motor B is locked)
“Error 18: Cooling circuit ‘A’ abnormal” Cooling circuit A is abnormal.
“Error 19: Cooling circuit ‘B’ abnormal” Cooling circuit B is abnormal.
<Status codes>
Indication
Meaning
“Status_1 Ambient temp abnormal” Ambient temp. is lower than 0, or higher than +35
“Status_2 Power-supply abnormal” Power supply voltage is 15% lower than standard.
“Status_3 Cooling circuits Overload” Compressor keeps running over 5 days.
(Overloaded run)
<Battery check>
Indication
Meaning
“Please exchange a battery” Battery accumulation time (2.8years) expires.
<Fan motor check>
Indication
Meaning
“Exchange fan motor ‘A’” Accumulation time of Condensing fan motor A
(5.6years) expires.
“Exchange fan motor ‘B’” Accumulation time of Condensing fan motor B
(5.6years) expires.
“Exchange fan motors ‘A and ‘B’” Accumulation times of both Condensing fan motor A
and B (5.6years) expire.
<Alarms>
Indication
Meaning
“High Temp Warning 20XX/XX/XX XX:XX:XX” Chamber temp. is higher than high temp. alarm set
temp. (Year/Month/Date Time/Minute/Second)
“Low Temp Warning 20XX/XX/XX XX:XX:XX” Chamber temp. is lower than low temp. alarm set
temp. (Year/Month/Date Time/Minute/Second)
“Power failure Warning 20XX/XX/XX XX:XX:XX” Power failure has occurred, power switch is in off
Position, or power cord is disconnected.
(Year/Month/Date Time/Minute/Second)