- 31 -
(3) Compressor maximum running time
Either Compressor A or B keeps running for 90 minutes at maximum for compressor protection.
Operation sequence
Ex: In compressor overlapping run
When a unit reaches to the point where either compressor A or B runs solely, Compressor A should
be turned on. Compressor B should be turned off (Point in the picture below)
When unit cannot reach set temp +0.4 within 90 minutes, Compressor A should be turned off and
Compressor B should be turned on. (Point in the picture below)
If a unit reached to the point to start compressor overlapping run when running time of Compressor B
exceeds 90 minutes, Compressor A should be turned on. (Point in the picture below)
When unit reaches to the point where either compressor A or B runs solely, Compressor B should be
turned off. (Point in the picture below)
(4) Diagnosis to switch operation mode
a) When unit reaches set temp -1.5 during compressor overlapping run, diagnosis is performed to
determine if the unit can keep cooling performance using a single compressor and operation mode
changes to compressor alternate run.
b) When unit reaches set temp +3.0 during compressor alternate run, diagnosis is performed to
determine if the unit cannot keep cooling performance using a single compressor, and operation
mode is changed to compressor overlapping run.
Compressor overlapping run Compressor alternate run
ĺ Compressor alternate run ĺ Compressor overlapping run
㩷㩷㩷㩷㩷㩷㩷㩷㩷㩷㩷㩷
5GVVGOR
5GVVGOR
5KPINGEQORTWP
#NVGTPCVGEQORTWP
%QOR
%QOR
 OKP  OKP
QH
H
QP
QP
QP
QH
H
Ԙ
ԙ
Ԛ
ԛ
QHH
QH
HQP
ON
㪦㪥
Set temp Set temp
OFF OFF
Point to change
operation mode
Comp Comp
Comp Comp
on
on of
f
of
f
on
on
of
f
of
f
on
on
of
f
of
f
of
f
on
Point to change
operation mode