EasyManua.ls Logo

Sanyo MDF-U500VX - Temperature Uniformity Data (Normal Mode); Uniformity Graphs at -80°C (Normal Mode)

Sanyo MDF-U500VX
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 51 -
Note) Following data are the reference only, so they does not assure product's performance.
For the graphs of ECO-mode, it does not assure re-production in actually.
<Conditions>
Ambient tem
p
: 20/30
Normal mode, ECO-mode
No load
Set temp: -80, Ambient temp:20, Normal mode, Air temp.
Unit:
Max. Min.
Center o
f
c
y
cle
Difference Max. Min.
Center o
f
c
y
cle
Difference
Upper Left front -77.5 -81.1 -79.3 ±1.8 -77.4 -81.2 -79.3 ±1.9
Left back -79.7 -84.6 -82.2 ±2.5 -79.6 -84.6 -82.1 ±2.5
Right front -78.9 -83.0 -81.0 ±2.1 -78.8 -83.2 -81.0 ±2.2
Right back -77.8 -81.4 -79.6 ±1.8 -77.7 -81.4 -79.6 ±1.9
Center
-80.0
-84.3 -82.2 ±2.2 -80.0 -84.3 -82.2 ±2.2
Middle Left front -77.9 -79.6 -78.8 ±0.8 -77.8 -79.7 -78.8 ±1.0
Left back -80.4 -82.7 -81.6 ±1.2 -80.4 -82.7 -81.6 ±1.2
Right front -79.3 -80.7 -80.0 ±0.7 -79.4 -80.6 -80.0 ±0.6
Right back -78.0 -79.9 -79.0 ±1.0 -77.9 -79.9 -78.9 ±1.0
Center
-80.5
-82.7 -81.6 ±1.1 -80.4 -82.7 -81.6 ±1.2
Lower Left front -76.6 -79.6 -78.1 ±1.5 -76.5 -79.6 -78.1 ±1.6
Left back -78.0 -83.1 -80.6 ±2.6 -78.0 -83.1 -80.6 ±2.6
Right front -78.3 -82.2 -80.3 ±2.0 -78.3 -82.1 -80.2 ±1.9
㽿
Right back -76.5 -80.2 -78.4 ±1.9 -76.6 -80.2 -78.4 ±1.8
Center -76.9 -81.9 -79.4 ±2.5 -76.8 -81.9 -79.4 ±2.6
Center of upper shelf
-78.9 -80.9 -79.9 ±1.0 -78.9 -81.0 -80.0 ±1.1
Center of lower shel
f
-78.3 -80.1 -79.2 ±0.9 -78.3 -80.2 -79.3 ±1.0
Average - - -80.0 - - - -80.0 -
AT 20, Normal mode
50Hz 60Hz
SV-80, AT20, Normal mode, 50Hz
㪄㪏㪋
㪄㪏㪊
㪄㪏㪉
㪄㪏㪈
㪄㪏㪇
㪄㪎㪐
㪄㪎㪏
㪄㪎㪎
㪄㪎㪍
㪄㪎㪌
㪄㪎㪋
㪇㪅㪇 㪇㪅㪌 㪈㪅㪇 㪈㪅㪌 㪉㪅㪇 㪉㪅㪌
Time[Hour]
Chamber temp.[
]
SV-80, AT20, Normal mode, 60Hz
㪄㪏㪋
㪄㪏㪊
㪄㪏㪉
㪄㪏㪈
㪄㪏㪇
㪄㪎㪐
㪄㪎㪏
㪄㪎㪎
㪄㪎㪍
㪄㪎㪌
㪄㪎㪋
㪇㪅㪇 㪇㪅㪌 㪈㪅㪇 㪈㪅㪌 㪉㪅㪇 㪉㪅㪌
Time[Hour]
Chamber temp.[
]

Related product manuals