RUNNING OPERATION (MENU/Set)
Chamber temperature setting (Temperature)
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is -35
o
C.
Note: The unit is set at the factory that the chamber temperature -80
o
C. (Setting range: -50 - -90
o
C)
Table 1. Basic operation sequence (Chamber temperature -80
o
C to -75
o
C)
Description of operation Key operated Indication after operation
1 Switch on the freezer. ----- The basic screen displayed
2 Press the MENU key. MENU
The menu window is indicated. And the
“set” is highlighted.
3 Press the ENTER key. ENTER
The temp. Setting window indicated. And
Chamber temperature -80 is highlighted.
4
The number input key is pressed in
order of 7 ĺ 5.
Chamber temperature is changed from -80
to -75.
5 Press the MENU key. MENU
The menu window is indicated. And the
“OK” is highlighted.
6 Press the ENTER key. ENTER
Chamber temperature is memorized, and
the top screen window is indicated.
Note:
The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation.
Although the value of the chamber temperature setting can range between -50
o
C and -90
o
C, the
guaranteed temperature without load is -85
o
C at ambient temperature of 30
o
C