VARIOUS SETTING (MENU/Tools)
Change of compressor delay time (Tools/Comp. Delay)
There is a starting delay time setup (Comp Delay) of A compressor /B compressor in the power failure
recovery in this product. That can be set up until 3-15 minutes. A compressor setup in the factory
shipment is 3 minutes. An A compressor starts first. (it is seen from the rear, the left)
A compressor starts B compressor, and starts after about one minute.
Table 13 Compressor starting delay time setup process (When the starting delay time of the
compressor is changed from 3 minutes to 5 minutes.)
Description of operation Key operated Indication after operation
1 -----
The top screen displayed
2 Press the MENU key. MENU
The menu window is indicated. And the
“set” is highlighted.
3 Press the ENTER key.
The temp. setting window indicated.
4
Select Date Time item pressing Shift
keys.
A menu window is indicated, and the log of Date
Time
is highlighted.
5
Select Ring Back item pressing
MENU key. Then, pressing ENTER
key
The Date Time setting window indicated.
6
Select Comp Delay item pressing
Shift keys.
Comp Delay highlights.
7
The number input key is pressed in
order of 0ĺ5.
The numerical value of the Comp Delay
(Comp Delay is changed at 3 to 5).
8
Press the MENU key.
MENU
The menu window is indicated. And the
“OK” is highlighted.
9
Press the ENTER key. ENTER
Date time is memorized, and the top screen
window is indicated.
ENTER