EasyManua.ls Logo

Sanyo MDF-U537D - W I R I N G D I a G R a; MDF-U333 Temperature Uniformity and Fluctuation

Sanyo MDF-U537D
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MDF-U333
Temperature uniformity and fluctuation
Test condition
Mains : 1φ100V50/60Hz
AT : 35
SV -25℃・-20℃・-30
No load Measure with type T sensor
Measuring pointCenter point of each container
(1).SV=-25℃
Temperature(
Fluctuation(
Top -21.8
-23.2
Middle -24.3 ± 2.0
-24.6
Bottom -24.2
Uniformity Max./Min -21.8/-24.6
Range(deg) 2.8(±1.4)
(2).SV=-20℃
Temperature(
Fluctuation(
Top -17.1
-18.3
Middle -19.2.6
-19.4
Bottom -19.0
Uniformity Max./Min -17.1/-19.4
Range(deg) 2.3(±1.15)
(3).SV=-30℃
Temperature(
Fluctuation(
Top -26.3
-28.1
Middle -29.3 ±1.7
-29.4
Bottom -29.1
Uniformity Max./Min -26.3/-29.4
Range(deg) 3.2(±1.6)
noteThese data are reference only.
Measuring point
1
2
3( 1/2 center air)
4
5
Measuring point
Measuring point
1
2
3( 1/2 center air)
4
5
5
1
2
3( 1/2 center air)
4
42−

Related product manuals