- 15 -
The procedure is explained as follow to calibrate temperature by turning VR03 on the PCB.
Step to function mode, F07 and adjust the difference between digital displayed temperature and
actual temperature by using a screwdriver.
<Ex.1>
In case digital displayed temperature is 100Υ and set temperature is 100Υ, while the actual
temperature reading is 99.74Υ:
э
Turn VR03 counterclockwise in 1 graduation.
 Temperature changes in every 0.26Υ graduation.
<Ex.2>
In case digital displayed temperature is 100Υ and SV is 100Υ, while the actual temperature
reading is 100.26Υ:
э
Turn VR03 clockwise in 1 graduation.
Temperature changes in every 0.26Υ graduation.
Note:
1) When you complete calibration, ensure to see how it will go.
2) It is recommended that thermometer which has high accuracy should be used.
3) Difference between digital displayed temperature and actual temperature would be
affected by internal temperature distribution or place where thermometer is located .
VR03
Turn VR03 by
screwdriver.
Temperature calibration