Put calibration weight on pan.
When display = calibration weight, remove the
weight, tare, replace weight on pan.
Display = calibration weight ±0.01g(±0.1g).
For calibration weight, see accessories.
For below-balance weighing, turn cover on the
bottom of the balance revealing below-balance
weighing device.
lf you get "CE" display, tare display and repeat
calibration procedure.