Specifications
Model B 410 B 610 B 3100P
Capacity/weighing range g 410 610 600/1,200/3,100
Readability g 0.01 0.01 0.01/0.02/0.05
Tare range (by subtraction) g -410 -610 -3,100
Standard deviation g
±0.005
±0.005
±0.01/0.01/0.025
Max. linearity g
±0.01
±0.01
±0.015/0.02/0.05
Stabilization time (typical) s 1.5 1.5 2
Display update s 0.1 – 0.8 (selectable)
Adaption to ambient conditions
and application requirements
by selection of one of 4 optimized filter levels
Stability range d 0.25 ... 64 (selectable)
Ambient temperature range K 273 – 313 (0
o
C ... +40
o
C)
Allowable relative humidity 15 … 85
(moisture-proof rating)
Sensitivity drift within
+10 … +30
o
C
/
o
C
±510
-6
Pan size
mm
in.
211 x 169
8.3 x 6.7
235 x 180
9.3 x 7.1
Balance housing (W x D x H)
mm
in.
220 x 235 x 60
8.7 x 9.3 x 2.4
250 x 270 x 65
9.8 x 10.6 x 2.6
Net weight kg/lbs. 4/8.8 4.5/8.8 5/11
Power requirements
(voltage + frequency)
115 or 230 V depending on the AC adapter used
50 – 60 Hz
Allowable voltage fluctuation -20% … +15%
Power consumption(typical) VA max. 7 7 7.5
Interface (optional)
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
transmissions rate: 150 ... 9,600 Baud