SQmin operation
Example Determining sample weights while monitoring the minimum sample
quantity
(in this example, SQmin = 100 g)
Configuration: The SQmin display must be turned on.
) t Place the container for the sample on the scale and tare.
t. Place the sample on the scale.
y The minimum sample quantity is not reached (symbol k).
p t Print the weight value.
| N + 90.0 ! |
t Place another sample on the scale.
y The minimum sample quantity is exceeded.
p t Print the weight value.
| N + 110.0 g |
k t Briefly press the k key to toggle between the measured value and
SQmin value.
y The value for the minimum sample quantity is displayed for four
seconds.