Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 digit* 3.6.1
2 digit 2 digits 3.6.2
5 digit 5 digits 3.6.3
10 digT. 10 digits 3.6.4
20 digT. 20 digits 3.6.5
50 digT. 50 digits 3.6.6
100 diG. 100 digits 3.6.7
200 diG. 200 digits 3.6.8
500 diG. 500 digits 3.6.9
1000 dI. 1000 digits 3.6.10
START Start of averaging 3.18.
manual Manual* 3.18.1
automat Automatic 3.18.2
ACTIVTY Animal activity 3.19.
0.1 perc. 0.1% of animal/object 3.19.1
0.2 perc. 0.2% of animal/object* 3.19.2
0.5 perc. 0.5% of animal/object 3.19.3
1 perc. 1% of animal/object 3.19.4
2 perc. 2% of animal/object 3.19.5
5 perc. 5% of animal/object 3.19.6
10 perc. 10% of animal/object 3.19.7
20 perc. 20% of animal/object 3.19.8
50 perc. 50% of animal/object 3.19.9
100 perc. 100% of animal/object 3.19.10
PRINT Autom. printout of results 3.20.
manual Off* 3.20.1
automat On 3.20.2
DIS.UNLD Static display of result after load removed 3.21.
CLEARED Display is fixed until unload
threshold reached* 3.21.1
PRESENT Fixed display until c is pressed 3.21.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform is
too light, the following will occur:
Error code Inf 29 appears,
The weighing platform is not initialized,
The preset reference sample quantity is saved.
Setting: Applic./Applic.1/ANIM.WG/Min.init menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The “digits" here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000
digits are required, you must place at least 1000 g (= 1000 intervals =1000 digits)
on the weighing platform for initialization.
Starting the Measurements The averaging routine does not begin until the fluctuation in weight value remains
below a defined threshold over three consecutive measurements. The tolerance limit
is defined as a percentage of the animal or object weight (for example, 0.1%, 0.2%,
64 Operating Instructions Combics Indicators
Operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CAISL1 and is the answer not in the manual?

Sartorius CAISL1 Specifications

General IconGeneral
BrandSartorius
ModelCAISL1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals