Entering the reference percentage value
The reference weight for 100% is entered using the keypad and the O key is 
pressed to initialize the application.
Reading the reference percentage value
The reference weight can be read using a barcode scanner. 
 
h
  The entered value remains active until deleted by pressing the c key or until 
overwritten by a new value. It remains saved after the scale is switched off. 
  Preparation  t  Open the APPLIC. / APPLIC.1 / PERCENT menu
  Available parameter settings   * = Factory setting
MIN.INIT Minimum load for initialization   3.6
  1 digit  1 digit*  3.6.1
  2 digit  2 digits  3.6.2
  5 digit  5 digits  3.6.3
  10 digT.  10 digits  3.6.4
  20 digT.  20 digits  3.6.5
  50 digT.  50 digits  3.6.6
  100 diG.  100 digits  3.6.7
  200 diG.  200 digits  3.6.8
  500 diG.  500 digits  3.6.9
  1000 dI.  1000 digits  3.6.10
RESOLUT Resolution for calculation of reference value   3.9. 
  DISP.ACC  Display accuracy  3.9.1*
  10 FOLD  Display accuracy + 1 decimal place  3.9.2
  100FOLD  Display accuracy + 2 decimal places  3.9.3
DEC.PLCS Decimal places in displayed result  3.10
  NONE   None  3.10.1*
  1 DEC.PL  1 decimal place  3.10.2
   2 DEC.PL  2 decimal places  3.10.3
  3 DEC.PL  3 decimal places  3.10.4
SAVE WT. Parameter for saving weight values   3.11
  STABIL.  With stability  3.11.1*
  ACC.STAB  With increased stability  3.11.2
REF.WP Reference weighing instrument   3.13.
  NO WP  No weighing platform selected  3.13.1*
  wp  1  Weighing platform WP1  3.13.2
  wp  2  Weighing platform WP2  3.13.3
CALC.DIS Display of calculated value  3.15. 
  RESID.QT  Residual quantity  3.15.1*
  LOSS   Loss  3.15.2
  )    t   To save the setting, press the ) key.
  ( ( ...  t   To exit setup: Press the ( key several times.
  Minimum load for initialization  You can set the minimum load here, i.e. the load that must be placed on the 
weighing platform in order to carry out the application. If the load on the platform 
is too light, the following will occur:
–  Error code Inf 29 appears,
–  The weighing platform is not initialized,
–  The preset reference sample quantity is saved.
68  Operating Instructions Combics Indicators
  Operation